6725 Semi-micro Calorimeter Instruction Manual
10-1
TROUBLESHOOTING
Error List
The calorimeter will run a number of diagnostic checks upon itself and will advise the
operator if it detects any error conditions. Most of these errors and reports will be self-
explanatory. The following list contains errors that are not necessarily self-evident and
suggestions for correcting the error condition.
A Misfire Condition Has Been Detected
This error will be generated in the event the
total temperature rise fails to exceed 0.5°C after the first minute of the post-period.
Fuse wire not intact or improperly formed
Breakdown of the insulator and o-ring on the insulated electrode assembly
Bomb not filled with oxygen
A bomb leak.
Poor bucket stirring
Ignition lead wires have broken internal strands
A Preperiod Timeout Has Occurred.
The calorimeter has failed to establish an
acceptable initial temperature, prior to firing the bomb, within the time allowed. Possible
causes for this error are listed below:
A bomb leak
Poor bucket stirring
Lid not tight
Initial bucket temperature outside the acceptable range (1-2 deg C below the
jacket setpoint)
The Current Run Has Aborted Due To Timeout.
The calorimeter has failed to establish
an acceptable final temperature within the time allowed. Possible causes for this error
are listed below:
A bomb leak.
Poor bucket stirring.
There Is A Problem With The Bucket Thermistor.
Possible electrical open or short. This
errors will result if the temperature probe response is not within the expected range.
Probe substitution can be useful in determining the cause of the problem (probe or
electronics). The valid working range of the probe resistance is 1000 to 5000 ohms.
Check connection to board.
Check quick disconnect.
Replace probe.
There Is A Problem With The Jacket Thermistor. Possible electrical open or short. This
errors will result if the temperature probe response is not within the expected range.
Vista de pagina 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 75 76

Comentarios a estos manuales

Sin comentarios

Philips AVENT SCF860 manuály

Uživatelské manuály a uživatelské příručky pro Mixéry Philips AVENT SCF860.
Poskytujeme 2 manuály pdf Philips AVENT SCF860 ke stažení zdarma podle typů dokumentů: Uživatelský manuál






Další produkty a příručky pro Mixéry Philips

Modely Typ Dokumentu
HR1741 Uživatelský manuál       Philips HR1741 User Manual, 100 stránky
HR7740 Uživatelský manuál       Philips HR7740 User Manual, 116 stránky
AVENT SCF860/20 Uživatelský manuál       Philips AVENT SCF860/20 User Manual, 108 stránky
HR7775 Uživatelský manuál   Philips HR7775 User Manual, 2 stránky
RI1760 Uživatelský manuál   Philips RI1760 User Manual, 2 stránky
HR1744 Uživatelský manuál   Philips HR1744 User Manual, 2 stránky
HR1843 Uživatelský manuál    Philips HR1843 Benutzerhandbuch, 48 stránky
HR1842 Uživatelský manuál       Philips HR1842 User Manual, 48 stránky
RI1774 Uživatelský manuál   Philips RI1774 User Manual, 2 stránky
HR7620 Uživatelský manuál       Philips HR7620 User Manual [ru] , 6 stránky
HR1371 Uživatelský manuál      Philips HR1371 User Manual, 156 stránky
Walita RI1764 Uživatelský manuál   Philips Walita RI1764 User Manual, 2 stránky
RI2035 Uživatelský manuál   Philips RI2035 User Manual, 2 stránky
HR3652/01 Uživatelský manuál       Philips HR3652/01 User Manual, 82 stránky