6400 Calorimeter Instruction Manual
5-17
Printer Port (RS232):
Accesses sub-menu, Printer Port Communications. Sets the communication parameters
for the RS232 ports used for the printer port. Standard options for data bits, parity, stop
bits, handshaking, baud rate and balance type are provided to match any devices that
might be connected to these ports.
Number of Data Bits.
Standard options for data
bits. Toggles between 7
and 8.
Parity. Standard options for
parity. Choose from None,
Odd or Even.
Number of Stop Bits.
Standard options for stop
bits. Toggles between 1
and 2.
Handshaking. Standard
options for handshaking.
Choose from Xon/Xoff,
RTS/CTS and None.
Baud Rate. Standard options for baud rate. Choose from
19.2K, 9600, 4800, 2400, 1800, 1200, 600, 300, 150, 134.5,
110, and 75.
Printer Type. Toggles between a Parr 1757 and a generic
printer. When set for the 1757 Printer, all of the features of
this printer, such as bold printing, will be activated.
Printer Port Loop Back Test. Used for factory testing of the
printer port.
Balance Port (RS232):
Accesses sub-menu, Balance Port
Communications.
Balance Type
Toggles through the
available balance templates.
Customize Balance Settings
Sets the communication
parameters for the RS232
port used for the balance
port. Standard options for
data bits, parity, stop bits,
handshaking, baud rate and balance type are provided to match any devices that
might be connected to these ports.
Number of Data Bits. Standard options for data bits. Toggles
between 7 and 8.