Flow Controller
www.parrinst.com
9
Navigating the Menus
To navigate to the menu select mode, hold down the
Return key and press the Up Arrow key. To navigate
in the select mode press the Up Arrow or Down Ar-
row key and then the Return key to enter the chosen
mode.
Once in the desired mode press the Return key to
navigate to each parameter, use the Up Arrow or
Down Arrow to change the respective parameter
value. Pressing the Auto/Man key or pressing the
Return key will save the new value.
Configuration Mode - ConF
Note: There are software “locks” in place to change certain
parameters.
To enter the configuration mode, cLoc = 10
*Remember to press the AUTO/MAN key to save
any new value.
Operation
Mode
Select
type/value
Comment
InPt 0_5 0-5VDC
ruL * Upper range flow
value of MFC
rLL 0 Lower range flow
value of MFC
dPoS 0 Decimal position
CtYP SnGL Primary Control
CtrL rEv Reverse Acting
ALA1 nonE No Alarm 1
ALA2 nonE No Alarm 2
LAEn diSA
Inhi nonE
USE1 rEtS Retransmit SP output
tyP1 0_5 0 to 5Vdc output 1
ro1H * Upper range flow
value of MFC
ro1L 0 Lower range flow
value of MFC
diSP 1
CLoc 10 Configuration Mode
Lock Code
Setup Mode - SEtP
To enter setup mode, uLoc = 20
Operation
Mode
Select
type/value
Comment
FiLt 2.0
OFFS 0 PV offset value
PPLJ Current Primary Out-
put Power. Read only
Pb_P 0
diFP 0.1
SPuL ro1H value from Con-
figuration Mode
SPLL 0
APt diSA Auto Pre-tune
PoEn diSA Manual Control Select
SPr diSA Set point ramp shown
in operating mode
rP blank
SP Current setpoint value
from operation mode
SLoc 20 Setup Mode Lock Code
Automatic Tune Mode – Atun
To enter Tune mode, tLoc = 30
Operation
Mode
Select type/value
Ptun OFF
Stun OFF
tLoc 30
Operation Mode - OPtr
Press the Return key to navigate to the SP value
display. Press the Up Arrow or Down Arrow key to
change the SP value. Press the Auto/Man key to
store the SP value.