6200Calorimeter O
p
eratin
g
Instruction Manual
Figure 4- 1
6200 Factory Default Settings
Calorimeter Operation
Operating Mode Determination
Bomb Installed/EE 1/2400.0
Heater and Pump OFF
Operating Controls
Method of Operation Dynamic
Reporting Units Btu/lb
Use Spiking Correction OFF
“OTHER” Multiplier 4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s) 1200 S
LCD Contrast 30%
Print Error Messages ON
Language English
Spike Controls
Use Spiking OFF
Heat of Combustion of Spike 6318.4
Use Fixed Spike OFF
Weight of Fixed Spike 0.0
Prompt for Spike before Weight OFF
Program Info and Control
Date XX/XX/XXXX
Time XX:XX
Software and Hardware Info
Settings Protect OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User / Factory Settings
User Setup ID 00000001
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Calibration Data & Control
Calibration Run Limit 10
EE Max Std Deviation 0.0
Heat of Combustion
of Standard 6318.4
Bomb Service Interval 500
Use Bomb 1
Bomb 1 Through 4
EE Value 2400.0
Protected EE Value OFF
Thermochemical Corrections
Standardization
Fixed Fuse Correction ON 15.0
Fixed Acid Correction ON 10.0
Fixed Sulfur Correction ON 0.0
Determination
Fixed Fuse Correction ON 15.0
Fixed Acid Correction ON 10.0
Fixed Sulfur Correction OFF 0.0
Calculate Net Heat of Combustion OFF
Calculation Factors
Acid Value is Nitric Acid Only ON
Acid Multiplier 0.0709
Sulfur Value is Percent ON
Sulfur Multiplier 0.6238
Fuse Multiplier 1.0
Use Offset Correction (ISO) OFF
Offset Value 0.0
Fixed Hydrogen OFF 0.0
Hydrogen Multiplier 50.68
Dry Calculation OFF
Factory Default
Settings