4
Program Installation & Control
Parr Instrument Company
16
Calorimeter Operations
Operating Mode Determination
Bomb Installed/EE 1/2400.0
Operating Controls
Method of Operation Dynamic
Reporting Units BTU/lb
Use Spiking Correction OFF
OTHER” Multiplier 4.1868
Calibrate Touchscreen
LCD Backlight Timeout(s)
1200 s
LCD Backlight Intensity 70%
Print Error Messages ON
Language English
Spike Controls
Use Spiking OFF
Heat of Combustion of Spike 6318.4
Use Fixed Spike OFF
Weight of Fixed Spike 0.0
Prompt for Spike before Weight OFF
Program Information and Controls
Date & Time Settings
Software and Hardware Info
Settings Protect OFF
User/Factory Settings
Feature Key
Bomb Type Select
User Function Setup
Cold Restart
User/Factory Settings
User Setup ID 61-1108
Reload Factory Default Settings
Reload User Default Settings
Save User Default Settings
Calibration Data & Controls
Calibration Run Limit
10
EE Max Std Deviation 0.0
Heat of Combustion of Standard 6318.4
Bomb Service Interval 500
Control Chart Parameters
Use Bomb 1
Bomb 1 Through 4
EE Value 2400.0
Protected EE Value OFF
Thermochemical Corrections Standardization
Fixed Fuse Correction ON 50
Acid Correction Fixed HNO
3
10.0
Fixed Sulfur Correction ON 0.0
Heat of Formation Sulfuric Acid
36.1
Heat of Formation Nitric Acid 14.1
Determination
Fixed Fuse Correction ON 50
Acid Correction Fixed HNO
3
10.0
Fixed Sulfur Correction OFF 0.0
Calculation Factors
Nitric Acid Factor 1.58
Acid Multiplier 0.0709
Sulfur Value is Percent ON
Sulfur Multiplier 0.6238
Fuse Multiplier 1.0
Use Offset Correction (ISO) OFF
Offset Value 0.0
Table 4-1
Factory Default Settings